Full description
The Bruker Biospec MRI scanner consists of a 9.4 Tesla, 300 mm ultrashield refrigerated 30 cm magnet interfaced to a scalable Bruker Avance III spectrometer that allows unparalleled flexibility for MRI scanning of small and medium sized research animals and other samples.
The system incorporates a cryoprobe, which enables maximum signal and image quality to be achieved for rodent mouse brain imaging. The system also includes 8 parallel RF transmit and receive channels for multi-coil phased array imaging and numerous coils to investigate nuclei other than standard protons or "water" imaging, notably carbon-13 coils to investigate metabolism, sodium-23 coils to investigate cell viability, phosphorus-31 coils to investigate tissue energetics and fluorine-19 coils for labelling of targeted tracers.
This system offers new potential for advancing preclinical research and molecular imaging in national and international researchers collaboration within at the University of Queensland and is available for use through the Australian National Imaging Facility: www.anif.org.au. The system was funded with assistance from the Queensland State Government under the Australian Commonwealth Government's Education Investment Fund scheme.
SPECTROMETER
UQ Bruker 9.4T, 30cm bore
Hardware = AVANCE III
IPSO Controller
Gradient Supervisor Unit tty04
Automatic hardware identification (HWIDS) PN=H14176, SN=0217
Parallel Transmit (PEX/MUX) PN=H12592, SN=0015
GRADIENT
DPP Preemphasis, GRASI Control Unit Interface
Gradient Coils
BGA20S (128,200 Hz/cm), with XYZ shims
BGA12S (285,000 Hz/cm, BFR W3307165/0048) with XYZ shims
Gradient Amplifiers
3 x IECO gradient amplifiers (IECO 300A 500V, a.k.a. IECO_400_700)
SHIM
TOSI interface tty01
Shim Power Supply
Bruker Shim Power Supply (BSPS 66010), Shims Z0, 2XY, YZ, X2-Y2, XZ, Z2
Max current = 10A and Max Voltage = 60.0V for each shim coil
6 channels
Total Amps = 25.0A
RF (transmit)
Bruker linear amplifiers
- BLA1000 (1 kW) IE 15 – 400MHz, PN=W1345501, SN=0284
- BLA1000 (1 kW) IE 15 – 400MHz, PN=W1345524, SN=0023
- BLA2H500 (500 W) 400 – 500 MHz PN=W1345508, SN=0034
- BLA2H500 (500 W) 400 – 500 MHz PN=W1345508, SN=0032
- BLA2H500 (500 W) 400 – 500 MHz PN=W1345508, SN=0033
- BLA2H500 (500 W) 400 – 500 MHz PN=W1345508, SN=0031
Bruker Preamplifier
- HPPR Cover module PN=Z103945, SN=00283, ECL=01.01
- Preamplifier 1 HPLNA BB31P
- Preamplifier 1 HPLNA 19F1H
- Preamplifier 1 HPLNA 19F1H
- Preamplifier 1 HPLNA 19F1H
- Preamplifier 1 HPLNA 19F1H
Signal generation, reception and digitisation
AQS
DRU-E Master Unit 1 PN=Z102520, SN=02977
RXAD400 PN=Z102116, SN=2747 ECL=05.01
DRU-E PN=Z102520, SN=02938
RXAD400 PN=Z102116, SN=2724, ECL=05.01
SGU/2 400 PN=Z103080 SN=03581 ECL=02.07
SGU/2 400 PN=Z103080 SN=03576 ECL=02.07
SGU/2 400 PN=Z103080 SN=03579 ECL=02.07
Pulse Splitter PN=Z104431 SN=00200 ECL=01.00
AQS2M - 1
DRU-E Master Unit 2 PN=Z102520 SN=02978
RXAD PN=Z102116 SN=2745 ECL=05.01
DRU-E PN=Z102520, SN=02980
RXAD400 PN=Z102116, SN=2744, ECL=05.01
DRU-E PN=Z102520, SN=02979
RXAD400 PN=Z102116, SN=2727, ECL=05.01
DRU-E PN=Z102520, SN=02936
RXAD400 PN=Z102116, SN=2726, ECL=05.01
DRU-E PN=Z102520, SN=02937
RXAD400 PN=Z102116, SN=2725, ECL=05.01
DRU-E PN=Z102520, SN=02931
RXAD400 PN=Z102116, SN=2746, ECL=05.01
RF Splitter PN=Z104432 SN=00207 ECL=02.01
AQS2M-2
SGU/2 400 PN=Z103080 SN=03603 ECL=02.07
SGU/2 400 PN=Z103080 SN=03620 ECL=02.07
SGU/2 400 PN=Z103080 SN=03623 ECL=02.07
SGU/2 400 PN=Z103080 SN=03629 ECL=02.07
SGU/2 400 PN=Z103080 SN=03625 ECL=02.07
SGU/2 400 PN=Z103080 SN=03633 ECL=02.07
SGU/2 400 PN=Z103080 SN=03575 ECL=02.07
RF Connection
RF cable connections (detected by 'confamp')
--------------------------------------------------------------------
SGU1 NORM output -> input 1 of transmitter 2 (BLA1000 IE 15-400MHZ INR W1345524/0023 at TCP/IP 149.236.99.245)
SGU1 AUX output -> open
SGU2 NORM output -> input 1 of transmitter 1 (BLA1000 IE 15-400MHZ W1345501/0284 at TCP/IP 149.236.99.246)
SGU2 AUX output -> input 2 of transmitter 2 (BLA1000 IE 15-400MHZ INR W1345524/0023 at TCP/IP 149.236.99.245)
SGU3 NORM output -> input 1 of transmitter 4 (BLA2H500 E 400-500MHZ W1345508/0032 at TCP/IP 149.236.99.242)
SGU3 AUX output -> open
SGU4 NORM output -> input 2 of transmitter 4 (BLA2H500 E 400-500MHZ W1345508/0032 at TCP/IP 149.236.99.242)
SGU4 AUX output -> open
SGU5 NORM output -> input 1 of transmitter 3 (BLA2H500 E 400-500MHZ W1345508/0034 at TCP/IP 149.236.99.243)
SGU5 AUX output -> open
SGU6 NORM output -> input 2 of transmitter 3 (BLA2H500 E 400-500MHZ W1345508/0034 at TCP/IP 149.236.99.243)
SGU6 AUX output -> open
SGU7 NORM output -> input 1 of transmitter 5 (BLA2H500 E 400-500MHZ W1345508/0033 at TCP/IP 149.236.99.241)
SGU7 AUX output -> open
SGU8 NORM output -> input 2 of transmitter 5 (BLA2H500 E 400-500MHZ W1345508/0033 at TCP/IP 149.236.99.241)
SGU8 AUX output -> open
SGU9 NORM output -> input 1 of transmitter 6 (BLA2H500 E 400-500MHZ W1345508/0031 at TCP/IP 149.236.99.240)
SGU9 AUX output -> open
SGU10 NORM output -> input 2 of transmitter 6 (BLA2H500 E 400-500MHZ W1345508/0031 at TCP/IP 149.236.99.240)
SGU10 AUX output -> open
Blanking cable connections (detected by 'confamp')
--------------------------------------------------------------------
transmitter 1 (BLA1000 IE 15-400MHZ W1345501/0284 at TCP/IP 149.236.99.246)
amplifier B-1000W uses blanking 2
transmitter 1 (BLA1000 IE 15-400MHZ W1345501/0284 at TCP/IP 149.236.99.246)
amplifier B-300W uses blanking 2
transmitter 2 (BLA1000 IE 15-400MHZ INR W1345524/0023 at TCP/IP 149.236.99.245)
amplifier B-1000W uses blanking 1
transmitter 2 (BLA1000 IE 15-400MHZ INR W1345524/0023 at TCP/IP 149.236.99.245)
amplifier B-300W uses blanking 1
transmitter 3 (BLA2H500 E 400-500MHZ W1345508/0034 at TCP/IP 149.236.99.243)
amplifier 1H-500W uses blanking 6
transmitter 3 (BLA2H500 E 400-500MHZ W1345508/0034 at TCP/IP 149.236.99.243)
amplifier 1H-500W uses blanking 7
transmitter 4 (BLA2H500 E 400-500MHZ W1345508/0032 at TCP/IP 149.236.99.242)
amplifier 1H-500W uses blanking 4
transmitter 4 (BLA2H500 E 400-500MHZ W1345508/0032 at TCP/IP 149.236.99.242)
amplifier 1H-500W uses blanking 5
transmitter 5 (BLA2H500 E 400-500MHZ W1345508/0033 at TCP/IP 149.236.99.241)
amplifier 1H-500W uses blanking 8
transmitter 5 (BLA2H500 E 400-500MHZ W1345508/0033 at TCP/IP 149.236.99.241)
amplifier 1H-500W uses blanking 9
transmitter 6 (BLA2H500 E 400-500MHZ W1345508/0031 at TCP/IP 149.236.99.240)
amplifier 1H-500W uses blanking 10
transmitter 6 (BLA2H500 E 400-500MHZ W1345508/0031 at TCP/IP 149.236.99.240)
amplifier 1H-500W uses blanking 11
LO cable connections (detected by 'confrecsel')
--------------------------------------------------------------------
SGU1 -> Rf-Splitter input 1 (-> REC1)
SGU2 -> SGU3 -> Rf-Splitter input 2 (-> REC1-8)
Preamplifier connections (detected by 'confrecpre')
--------------------------------------------------------------------
Tune-SGU1 -> HPLNA BB31P -> REC1
Tune-SGU1 -> HPLNA 19F1H -> REC1
Tune-SGU1 -> HPLNA 19F1H -> REC2
Tune-SGU1 -> HPLNA 19F1H -> REC2
Tune-SGU1 -> HPLNA 19F1H -> REC2
User Contributed Tags
Login to tag this record with meaningful keywords to make it easier to discover
- Handle : 102.100.100/50043