K-axis dissolved iron (dFe) data from the Kerguelen-Axis region of the Southern Ocean

Australian Ocean Data Network
Schallenberg, C. and Bowie, A. ; SCHALLENBERG, CHRISTINA ; BOWIE, ANDREW ROSS
Viewed: [[ro.stat.viewed]] Cited: [[ro.stat.cited]] Accessed: [[ro.stat.accessed]]
ctx_ver=Z39.88-2004&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Adc&rfr_id=info%3Asid%2FANDS&rft_id=Dataset DOI&rft.title=K-axis dissolved iron (dFe) data from the Kerguelen-Axis region of the Southern Ocean&rft.identifier=Dataset DOI&rft.publisher=Australian Antarctic Data Centre&rft.description=Sampling was conducted according to GEOTRACES protocols. Samples for trace element analyses, including dissolved iron (dFe), were filtered through acid-cleaned 0.2 um cartridge filters (Pall Acropak) under constant airflow from several ISO class 5 HEPA units. All plastic ware was acid-cleaned prior to use, following GEOTRACES protocols. Samples were collected into low-density polyethylene (LDPE) bottles, acidified immediately to pH 1.7 with Seastar Baseline hydrochloric acid (HCl), double-bagged and stored at room temperature until analysis on shore. Samples for dFe analysis were pre-concentrated offline (factor 40) on a SeaFAST S2 pico (ESI, Elemental Scientific, USA) flow injection system with a Nobias Chelate-PA1 column. Samples were eluted from the column in 10% distilled nitric acid (HNO3), with calibration based on the method of standard additions in seawater (made using multi-element standards in a 10% HNO3 matrix, rather than an HCl matrix). Pre-concentrated samples were analysed using Sector Field Inductively Coupled Plasma Mass Spectrometry (SF-ICP-MS, Thermo Fisher Scientific, Inc.). Data were blank-corrected by subtracting an average acidified milli-Q blank that was treated similarly to the samples. The dFe detection limit for a given analysis run on the SeaFAST/SF-ICP-MS was calculated as 3 x standard deviation of the milli-Q blank on that run. Detection limits ranged from 0.016 to 0.067 nmol kg-1, with a median of 0.026 nmol kg-1 (n=12). GEOTRACES reference materials were analyzed along with samples and results were in good agreement with consensus values: SAFe D1 was measured at 0.69 +/- 0.05 nmol kg-1 (n=7; consensus value = 0.67 +/- 0.04 nmol kg-1) and GD was measured at 1.02 +/- 0.01 nmol kg-1 (n=6; consensus value = 1.00 +/- 0.1 nmol kg-1). Comments regarding the data spreadsheet: NaN = no sample dFe QC flags: 1 = high confidence in data quality 2 = detection limit 3 = low confidence in data quality detection limits: dFe data that were below the daily detection limit were replaced with the respective detection limit. They are flagged with the number 2 in the dFe QC flag column.Progress Code: completedStatement: No problems with analysis, but there were a lot of problems with GO-FLO bottles during sampling, i.e. bottles either not closing or leaking on arrival on deck. As a result, there is a large number of missing samples (NaN), and we also took some salinity samples for additional data quality control as follows: A subset of salinity samples (n=124) was taken from GO-FLO bottles to check whether bottles had closed at the expected depths and/or whether there were problems with leaking. Salinity samples were compared to CTD salinities, and samples where measured salinity did not match the respective CTD salinity (n=9) were flagged as suspect and are not included in the data sheet. Note that salinity checks were predominantly performed on GO-FLO bottles that were suspicious (e.g. leaking), so they do not represent a random set of samples.&rft.creator=Schallenberg, C. and Bowie, A. &rft.creator=SCHALLENBERG, CHRISTINA &rft.creator=BOWIE, ANDREW ROSS &rft.date=2018&rft.coverage=westlimit=70; southlimit=-66; eastlimit=95; northlimit=-57&rft.coverage=westlimit=70; southlimit=-66; eastlimit=95; northlimit=-57&rft_rights=This metadata record is publicly available.&rft_rights=These data are publicly available for download from the provided URL.&rft_rights= https://creativecommons.org/licenses/by/4.0/legalcode&rft_rights=This data set conforms to the CCBY Attribution License (http://creativecommons.org/licenses/by/4.0/). Please follow instructions listed in the citation reference provided at http://data.aad.gov.au/aadc/metadata/citation.cfm?entry_id=AAS_4344_dFe when using these data. http://creativecommons.org/licenses/by/4.0/).&rft_rights=Portable Network Graphic&rft_rights=https://i.creativecommons.org/l/by/3.0/88x31.png&rft_rights=Creative Commons by Attribution logo&rft_rights=Attribution 4.0 International (CC BY 4.0)&rft_rights=Legal code for Creative Commons by Attribution 4.0 International license&rft_rights=Attribution 4.0 International (CC BY 4.0)&rft_rights= https://creativecommons.org/licenses/by/4.0/legalcode&rft.type=dataset&rft.language=English Access the data

Licence & Rights:

Other view details
Unknown

https://creativecommons.org/licenses/by/4.0/legalcode

This data set conforms to the CCBY Attribution License (http://creativecommons.org/licenses/by/4.0/).

Please follow instructions listed in the citation reference provided at http://data.aad.gov.au/aadc/metadata/citation.cfm?entry_id=AAS_4344_dFe when using these data.
http://creativecommons.org/licenses/by/4.0/).

Attribution 4.0 International (CC BY 4.0)

https://creativecommons.org/licenses/by/4.0/legalcode

This metadata record is publicly available.

These data are publicly available for download from the provided URL.

Portable Network Graphic

https://i.creativecommons.org/l/by/3.0/88x31.png

Creative Commons by Attribution logo

Attribution 4.0 International (CC BY 4.0)

Legal code for Creative Commons by Attribution 4.0 International license

Access:

Other

Contact Information

metadata@aad.gov.au

Brief description

Sampling was conducted according to GEOTRACES protocols. Samples for trace element analyses, including dissolved iron (dFe), were filtered through acid-cleaned 0.2 um cartridge filters (Pall Acropak) under constant airflow from several ISO class 5 HEPA units. All plastic ware was acid-cleaned prior to use, following GEOTRACES protocols. Samples were collected into low-density polyethylene (LDPE) bottles, acidified immediately to pH 1.7 with Seastar Baseline hydrochloric acid (HCl), double-bagged and stored at room temperature until analysis on shore.

Samples for dFe analysis were pre-concentrated offline (factor 40) on a SeaFAST S2 pico (ESI, Elemental Scientific, USA) flow injection system with a Nobias Chelate-PA1 column. Samples were eluted from the column in 10% distilled nitric acid (HNO3), with calibration based on the method of standard additions in seawater (made using multi-element standards in a 10% HNO3 matrix, rather than an HCl matrix). Pre-concentrated samples were analysed using Sector Field Inductively Coupled Plasma Mass Spectrometry (SF-ICP-MS, Thermo Fisher Scientific, Inc.). Data were blank-corrected by subtracting an average acidified milli-Q blank that was treated similarly to the samples.

The dFe detection limit for a given analysis run on the SeaFAST/SF-ICP-MS was calculated as 3 x standard deviation of the milli-Q blank on that run. Detection limits ranged from 0.016 to 0.067 nmol kg-1, with a median of 0.026 nmol kg-1 (n=12). GEOTRACES reference materials were analyzed along with samples and results were in good agreement with consensus values: SAFe D1 was measured at 0.69 +/- 0.05 nmol kg-1 (n=7; consensus value = 0.67 +/- 0.04 nmol kg-1) and GD was measured at 1.02 +/- 0.01 nmol kg-1 (n=6; consensus value = 1.00 +/- 0.1 nmol kg-1).

Comments regarding the data spreadsheet:
NaN = no sample
dFe QC flags: 1 = high confidence in data quality
2 = detection limit
3 = low confidence in data quality
detection limits: dFe data that were below the daily detection limit were replaced with the respective detection limit. They are flagged with the number 2 in the dFe QC flag column.

Lineage

Progress Code: completed
Statement: No problems with analysis, but there were a lot of problems with GO-FLO bottles during sampling, i.e. bottles either not closing or leaking on arrival on deck. As a result, there is a large number of missing samples (NaN), and we also took some salinity samples for additional data quality control as follows:

A subset of salinity samples (n=124) was taken from GO-FLO bottles to check whether bottles had closed at the expected depths and/or whether there were problems with leaking. Salinity samples were compared to CTD salinities, and samples where measured salinity did not match the respective CTD salinity (n=9) were flagged as suspect and are not included in the data sheet. Note that salinity checks were predominantly performed on GO-FLO bottles that were suspicious (e.g. leaking), so they do not represent a random set of samples.

Notes

Purpose
Supply dissolved iron data for the K-axis region. Complimentary to other K-axis data.

Data time period: 2017-01-10 to 2018-02-28

95,-57 95,-66 70,-66 70,-57 95,-57

82.5,-61.5

text: westlimit=70; southlimit=-66; eastlimit=95; northlimit=-57